Evaluating commercially available wireless cardiovascular monitors for measuring and transmitting real-time physiological responses in children with autism.

Heather J Nuske, Matthew S Goodwin, Yelena Kushleyeva and 10 others

PMID 34741438

WHAT IT FOUND

In 55 children, two inexpensive heart-rate monitors stayed on and recorded usable data during rest and stress.

They showed higher heart rate and lower variability under stress. This is device testing, not treatment evidence.

Key findings

01Two of the three monitors tested with children were kept on by at least 80% of each group for the full session, so the third was dropped from further testing.

02The Mio Fuse wristband met the data-quality criteria for 96% of participants at rest and 97% during stress; the Polar H7 chest strap met them for 83% at rest and 96% during stress.

03During the stress task, both monitors recorded higher heart rate and lower heart-rate variability than during rest.

STILL TO COME

How it was doneWhat they foundWhat it means for OTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

Only 24 adults were used to screen the monitors, and only 55 children were tested in the main study, so the results are small and may not apply to all children with autism. Few children in the autism group needed very substantial support, and the study covered ages 8 to 12 only, so it says little about younger children or those with more severe support needs. The monitors were tested in a laboratory for about two hours, not in homes, schools, or everyday settings. Low-quality data often happened when children moved or adjusted the device, so active or unpredictable settings may reduce reliability. The study did not measure body mass index, blood pressure, anxiety, or depression, so differences in these factors could affect heart-rate readings. The devices' own heartbeat algorithms were not available, so the researchers could not independently check their accuracy. Because several monitors were worn at once, wireless interference between them could not be ruled out. The PulseOn wristband did not meet the 80% wear threshold and was not tested further, so its performance in children remains uncertain. The stress task was a laboratory toy-box delay, not a real-world stressor, so it may not match everyday triggers.

Declared interests

The authors declared no conflict of interest, and no monitor manufacturers were involved in conceptualization, design, implementation, analyses, or interpretation.

The easy way to misread this

Do not read this as evidence that heart-rate monitors can diagnose stress or guide therapy for children with autism. The study tested whether two devices stayed on, recorded usable data, and changed during a lab stress task; it did not test treatment benefit, and it did not include younger children or many children who need very substantial support.

Read it on PubMed →