WHAT IT FOUND

In hemiplegic stroke patients, the Honda device and a three minute assisted walking exercise did not immediately increase walking speed.

Hip angles changed, but the separate effects of the device and exercise cannot be separated.

Key findings

01The assisted trial used the Honda device and a three minute assisted walking exercise together, so the separate effect of the device cannot be isolated.

02During the assisted trial, hip flexion, extension on the paralyzed side and scissor angles increased, but walking speed, step length and cadence did not differ significantly.

03The device's walking measurements were repeatable in healthy participants.

STILL TO COME

How it was doneWhat they foundWhat it means for PTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

The stroke sample was small, with 16 patients, so the results are not enough to guide practice. The reliability test was done in healthy participants, and only 7 of the 10 recruited were analysed, so it does not show that the device measures stroke gait reliably. Three healthy participants were excluded because their walking time changed by 0.5 seconds or more, which may affect the reliability result. The assisted trial included both the device and a three minute assisted walking exercise, so the device alone cannot be identified as the cause of any change. The walking order was fixed as without, with, then without, so practice or fatigue effects cannot be ruled out. The study tested only immediate walking, not long term training or daily life function. The device only measured and assisted the hip joint, so the relationship with knee and ankle movement was not directly tested. Patients were divided into speed change groups after the test, so the toe grip and orthosis findings are associations, not proof of cause.

Declared interests

The paper reports no funding and no conflicts of interest directly relevant to the content.

The easy way to misread this

Do not conclude that the Honda device alone changed hip movement or improved walking. The assisted trial also included a three minute assisted walking exercise, and walking speed did not increase.

Read it on PubMed →