Motion-Steered Noise Management in Hearing Aids Reduces Listening Effort and Stress Indexed by Pupil Size and Heart Rate.
Jeppe H Christensen, Valentina Zapata-Rodríguez, Sébastien Santurette and 1 others
PMID 41975566WHAT IT FOUND
The two hearing aids produced the same speech-in-noise comprehension scores, but not the same effort.
In the harder condition with four competing talkers, the aid that also uses movement information kept pupil dilation and heart rate lower.
Key findings
01Comprehension fell significantly when four interfering talkers were present instead of two, and was the same with both hearing aids: 73.96% versus 76.56% correct with two talkers, and 63.02% versus 66.15% with four.
02In the harder four-talker condition, both pupil dilation and heart rate were significantly higher with hearing aid A than with hearing aid B, even though comprehension was equivalent.
03Self-reported listening effort and sound quality did not differ between the two hearing aids; only engagement did, and it was rated lower with hearing aid A.
STILL TO COME
How it was doneWhat they foundWhat it means for SLPs
Read the rest of this summary
You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.
What it does not show
Only 24 of the 32 participants had usable pupil recordings and only 20 had usable heart-rate recordings. The 12 dropped from the heart-rate analysis were removed because too little of the wristband signal was usable, which the authors link to hand movements carried over from the earlier task. The study was paid for by Oticon A/S, the company that makes both hearing aids tested, although the authors state the funder had no role in the design, conduct, analysis or write-up. This was one laboratory session. Nothing here shows whether the differences in pupil dilation and heart rate between the aids last beyond the session or matter to people in daily life. Self-rated effort was collected once after each block of 24 trials, so it could not capture effort changing from trial to trial, and it showed no difference between the aids even though the physiological measures did. The explanation offered for the difference between the aids, that hearing aid B adapts its processing using movement information, comes from how the two devices are built rather than from a test that switched that feature on and off in the same aid. The listening task was hard for everyone and the news clips lasted 33 seconds, which is much shorter than a real conversation.
Declared interests
Funded by Oticon A/S, which makes both hearing aids tested. The authors state the funder had no role in the design or conduct of the study, in collecting, analysing or interpreting the data, in preparing or approving the article, or in the decision to submit it. The authors declare no other conflicts of interest.
The easy way to misread this
Do not read this as evidence that motion-steered noise management improves hearing. The two aids produced the same speech comprehension scores, and the difference between them was in pupil dilation and heart rate during a single laboratory session, funded by the company that makes both devices.
Summarised by AI from the full paper, without a clinician reviewing it. Check it against the source before it changes what you do. Read it on PubMed →