PTRCTProsthetics and orthotics international2026

Evaluating modified socket casting and molding vs. total surface bearing techniques in unilateral transtibial amputation: Effects on rotational control, socket fit satisfaction, and interface pressure.

Batoul Bagheripour, Mohammad Ali Mardani, Behnam Hajiaghaei and 3 others

PMID 41452290

WHAT IT FOUND

Socket-liner interface pressure did not differ between modified and total surface bearing sockets.

Participants reported greater stability and preferred the modified socket, and it had less lateral rotation after walking.

Key findings

01Modified socket stability score was higher than TSB (81.25 vs 68.75, P = 0.02).

02Socket-liner interface pressure showed no significant difference between modified and TSB sockets (P > 0.05).

03After walking 3 m, lateral rotation was higher in the TSB socket than in the modified socket (P = 0.01).

STILL TO COME

How it was doneWhat they foundWhat it means for PTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

Only eight participants completed the trial, and two of ten dropped out. All participants were at K3 activity level, so the findings may not apply to other activity levels. Interface pressure was measured only in static weight-bearing positions and during walking, not across gait phases or on ramps, stairs, or uneven terrain. The rotation measurement method was not validated against gold-standard motion analysis. The study did not control for confounding variables such as muscle tonus and perspiration during static testing. The follow-up period was 4 weeks per socket, with no long-term assessment after limb volume changes. The prosthetist who fabricated and fitted the sockets was not blinded, although the assessor was. The sample size was based on interface pressure, and that outcome showed no significant difference.

Declared interests

The authors disclosed no financial support for the research, authorship, or publication.

The easy way to misread this

Do not conclude that the modified socket is proven better for transtibial amputation fitting. The interface pressure result was null, only eight completed, and the rotation method was not validated.

Summarised by AI from the full paper, without a clinician reviewing it. Check it against the source before it changes what you do. Read it on PubMed →