Antimicrobial Efficacy of a New Chlorhexidine-based Device Against Staphylococcus aureus Colonization of Venous Catheters.
Paulina M Kowalewska, Shawn M Petrik, Attilio E Di Fiore and 1 others
PMID 29489705WHAT IT FOUND
A chlorhexidine lock connector lowered Staphylococcus aureus bacterial counts in lab solutions and pig catheters, especially with saline.
Heparin or citrate needed longer contact. It was not tested in patients.
Key findings
01In 0.9% NaCl lock solution, ChloraLock reduced S aureus CFU by more than 6 log10 after 20 minutes, and no colonies grew.
02With 100 units/mL heparin or sodium citrate, 20 minutes did not significantly reduce CFU, but 4 hours produced approximately 4- to 5-log10 reductions.
03In pigs, 48 hours after catheter contamination, S aureus grew from 7 of 8 control lumens but 1 of 8 ChloraLock-treated lumens.
STILL TO COME
How it was doneWhat they foundWhat it means for RNs
Read the rest of this summary
You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.
What it does not show
The study tested lab solutions and pigs, not patients, so it does not show that ChloraLock prevents bloodstream infections in people. The device is investigational, and the authors say studies are needed to determine safety and feasibility in a patient population. Pig catheters were in place for only 7 days, so results apply only to short-term catheters. Only Staphylococcus aureus was tested, so it says nothing about other organisms that can colonize catheters. In pigs, clot formation meant lock fluid was not retrievable from 21% of lumens on average, so some samples could not be analysed. The effect of lock solution type was important: heparin and citrate needed longer contact than saline, and times between 20 minutes and 4 hours were not tested. Scanning electron microscopy examined only small 1-cm catheter segments, so biofilm and clot patterns may not represent the whole catheter.
The easy way to misread this
Do not conclude that ChloraLock prevents catheter-related bloodstream infections in patients. The study measured bacterial counts in lab solutions and pig catheters, not patient infections, and the device is investigational.