PTOTSLPOtherFolia phoniatrica et logopaedica : official organ of the International Association of Logopedics and Phoniatrics (IALP)2026

Accuracy of Balloons and Handheld Spirometers as Low-Cost Tools for Measuring Vital Capacity and Air Volume: A Laboratory-Based Validation Study.

James A Curtis, Mel Grasso, Brandon Jagdhar

PMID 41591963

WHAT IT FOUND

In a lab setting, a $150 digital spirometer and a $1 balloon measured injected air volumes with small, predictable errors compared to expensive clinical equipment.

These low-cost tools accurately estimate vital capacity when simple mathematical corrections are applied.

Key findings

01All four low-cost tools showed substantial agreement with reference standards, with the digital spirometer having the smallest median error (0.07 L vs ground truth; 0.15 L vs gold standard).

02Balloon measurements were less accurate than spirometers due to shape variability, but applying linear regression correction equations improved accuracy for all tools.

03The study used a calibration syringe to inject known volumes, meaning results reflect tool precision on static air, not patient effort.

STILL TO COME

How it was doneWhat they foundWhat it means for PTsWhat it means for OTsWhat it means for SLPs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

No human participants were involved; air was injected by a syringe, so the study does not account for errors caused by poor patient seal, weak effort, or inconsistent blowing technique. The results apply only to the specific brands and models tested (Hygloss balloon, Baseline analogue spirometer, Contec digital spirometer). Other brands may behave differently. The balloon methods rely on the assumption that an inflated balloon is a sphere, which is geometrically inaccurate, leading to greater variability at higher volumes. The correction equations are derived from this specific dataset and require external validation before being widely adopted in clinical practice.

Declared interests

The authors declared no conflicts of interest. The study received no sponsorship or funding.

The easy way to misread this

Do not assume these tools are accurate 'out of the box' for patients. The study measured static air injection, not human effort, and found systematic errors that required mathematical correction equations to fix. Without applying these corrections, low-cost tools may over- or under-estimate vital capacity.

Summarised by AI from the full paper, without a clinician reviewing it. Check it against the source before it changes what you do. Read it on PubMed →