A novel use of inertial sensors to measure the craniocervical flexion range of motion associated to the craniocervical flexion test: an observational study.
Tomás Pérez-Fernández, Susan Armijo-Olivo, Sonia Liébana and 4 others
PMID 33213452WHAT IT FOUND
Forehead inertial sensors measured neck flexion during the craniocervical flexion test reliably, but healthy young adults varied too much for standard range-of-motion targets at each pressure stage.
Key findings
01The forehead inertial sensor showed good or excellent reliability for measuring craniocervical flexion range of motion during each craniocervical flexion test stage in 56 asymptomatic young adults.
02Craniocervical flexion range of motion increased progressively with each pressure stage, but between-person variability was large enough that standard range-of-motion targets for each stage were not supported.
03The findings apply to asymptomatic young university adults, not to patients with neck pain, headache, or temporomandibular disorders.
STILL TO COME
How it was doneWhat they foundWhat it means for PTs
Read the rest of this summary
You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.
What it does not show
Only asymptomatic, young university adults were tested, so the findings do not show how patients with neck pain, headache, temporomandibular disorders, or older adults would perform. The study did not use electromyography to confirm whether superficial neck muscles were compensating, so some participants may have used incorrect strategies. Between-person variability was large, and the standard error of measurement was too high relative to the small differences between consecutive pressure stages, so standard range-of-motion targets could not be set. The sensor was tested only during the supine craniocervical flexion test, not during functional postures or as a treatment outcome measure. The paper states 56 participants but reports 18 males and 23 females, so the sex breakdown is unclear.
The easy way to misread this
Do not read the reliability result as proof that the sensor improves craniocervical flexion test performance or neck pain treatment. The study measured only 56 asymptomatic young adults, did not test patients, and found too much between-person variability to set standard range-of-motion targets for each pressure stage.