PTOtherJournal of rehabilitation medicine2026

Validation of heart rate measurements from the polar verity sense during the submaximal Ekblom-Bak test in adult individuals with obesity.

Grete Rudi Onstad, Fredrik Klæboe Lohne, Marius Steiro Fimland

PMID 42702963

WHAT IT FOUND

In adults with obesity, arm-worn Polar Verity Sense matched chest strap heart rate during steady-state cycling, giving nearly identical estimated VO2 max.

Most participants preferred the arm-worn device, but it was tested only during cycle ergometry.

Key findings

01The two devices gave almost the same values: agreement coefficients were 0.998 to 0.999, and average percentage error was 0.46% to 0.51%.

02The estimated VO2 max values from the arm-worn device were almost identical to the chest strap: average differences were -0.025 ml/kg/min for forearm placement and -0.016 ml/kg/min for upper-arm placement.

03Twenty-nine participants preferred the upper-arm device, and 26 participants placed both arm-worn devices but needed help with the chest strap.

STILL TO COME

How it was doneWhat they foundWhat it means for PTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

Participants were a convenience sample from one inpatient obesity rehabilitation programme, and no a priori sample size calculation was done for the validation analysis. The criterion device was the Polar H10 chest strap, not an ECG, because ECG recording was not available at the facility. The test only examined steady-state cycling at low and higher work rates, so the results do not show how the device performs during fluctuating intensity or other activities. Estimated VO2 max agreement was not an independent validation of VO2 max, because VO2 max was calculated from heart rate. Participants who had previously done the test kept their earlier workload, which reduced individualization and was reflected in one participant being unable to complete the test. The questionnaire responses were analysed descriptively, so preference and discomfort findings are participant-reported experiences rather than statistically analysed differences. The upper-arm device had a single size, and 2 participants reported discomfort from tightness despite maximum strap length. The validity analyses used one averaged heart rate value per participant at each work rate, so the number of paired comparisons was lower than the total number of second-by-second heart rate recordings.

Declared interests

The authors declared no conflicts of interest. The study was internally funded by NTNU.

The easy way to misread this

Do not conclude that the Verity Sense independently measures VO2 max, works during fluctuating exercise, or applies outside adults with obesity in this cycle test. The study tested steady-state cycling only, and VO2 max estimates were calculated from heart rate rather than measured directly.

Read it on PubMed →


The study

Participants
49 participants for estimated VO2 max analyses; 48 for steady-state heart rate analyses; 51 for feasibility analyses
Certainty of evidence
Low

More in cardiopulmonary

Browse

Cite

Grete Rudi Onstad, Fredrik Klæboe Lohne, Marius Steiro Fimland Validation of heart rate measurements from the polar verity sense during the submaximal Ekblom-Bak test in adult individuals with obesity. Journal of rehabilitation medicine. 2026.

Read the original