Validation of a Portable Wireless Force Platform System To Measure Ground Reaction Forces During Various Tasks.

Vasileios Mylonas, Christos Chalitsios, Thomas Nikodelis

PMID 38050544

WHAT IT FOUND

Portable wireless force plates showed high agreement with laboratory gold standards for measuring vertical force during standing and jumping in three participants.

Deviations from known weights were under 0.1 kg. Results suggest these devices are valid alternatives for clinical and sports use.

Key findings

01Deviations from known masses ranged from 0.01 ± 0.04 kg to 0.04 ± 0.03 kg across weights up to 82.18 kg.

02Bland-Altman analysis showed limits of agreement for jump impulse ranging from approximately -0.52 to 5.54 N⋅s.

03The portable plates are 1-D, limiting their use for multidimensional analysis like joint moments during gait.

STILL TO COME

How it was doneWhat they foundWhat it means for PTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

Only three participants were tested, all of whom were physical education students familiar with the tasks. The portable plates are one-dimensional, so they cannot measure multidimensional forces required for gait analysis. The plates were not permanently fixed to the surface, creating a risk of slippage during dynamic jumps.

Declared interests

No specific funding or conflict of interest declarations are provided in the text.

The easy way to misread this

Do not assume these plates can replace laboratory systems for all biomechanical analyses. They measure only vertical force, so they cannot provide the multidimensional data needed for gait or joint moment analysis.