Convergent validity of a simplified device and relationship between blood lactate and salivary lactate after a vertical squat jump in healthy non-athletes.

Satomi Okano, Hitomi Nishizawa, Joya Yui and 3 others

PMID 33814702

WHAT IT FOUND

A portable saliva lactate device followed the same pattern as a lab analyzer in four participants, but after a vertical squat jump, blood and saliva lactate did not track together at the same times.

Saliva lagged by 5 minutes.

Key findings

01In the 4 participants analysed, normalized saliva lactate values from Lactate Pro 2 showed overall agreement with the automatic analyzer of ICC 0.773 (95% confidence interval 0.678 to 0.844).

02In the 15 participants analysed after the vertical squat jump, normalized blood lactate and saliva lactate showed low or negative agreement at the same time points, including ICC 0.012 immediately after and -0.646 at 30 min.

03Saliva lactate lagged behind blood lactate by 5 minutes after the vertical squat jump, with a cross correlation of 0.750 at lag 1.

STILL TO COME

How it was doneWhat they foundWhat it means for PTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

The study used healthy participants only, and the vertical squat jump test included only males, so it does not show what happens in patients or women. Only 4 participants were analysed in the device test and 15 in the vertical squat jump analysis, so the agreement estimates are based on a very small sample. Lactate Pro 2 cannot measure saliva lactate below 0.4 mmol/L or less; 'Lo' readings forced exclusion of 1 of 5 device-test participants and 5 of 20 vertical squat jump participants. The paper says all 20 vertical squat jump participants were analyzed, but it also says 5 were excluded from analysis and Table 4 reports 15 analysed. Absolute saliva values differed between Lactate Pro 2 and the automatic analyzer, so the comparison relied on normalized values rather than raw concentrations. The vertical squat jump was only one exercise task, and the authors note the saliva-blood time lag may depend on the exercise used. No clinical exercise prescription, patient outcome, or treatment effect was tested.

Declared interests

This study was supported by an Intramural Research Grant (26-6) for Neurological and Psychiatric Disorders of the National Center of Neurology and Psychiatry and a Grant-in-Aid for Scientific Research (19K12862 to H.N.). The authors declared no conflict of interest.

The easy way to misread this

Do not use saliva lactate from Lactate Pro 2 as a direct replacement for blood lactate after a vertical squat jump. The device could not measure baseline saliva in 5 of 20 participants, and blood and saliva lactate did not agree at the same time points.