Comparison of Common Methodologies for the Determination of Knee Flexor Muscle Strength.

Dan I Ogborn, Alix Bellemare, Brittany Bruinooge and 3 others

PMID 33842031

WHAT IT FOUND

Knee flexor strength measured by a wall-anchored handheld dynamometer correlates with lab equipment but cannot be swapped for it.

Seated testing yields higher torque than supine, and dorsiflexing the foot increases force compared to plantarflexion.

Key findings

01Values from a fixed handheld dynamometer and an isokinetic dynamometer are correlated but have wide limits of agreement, so they should not be interchanged.

02Isometric knee flexion torque is significantly greater when the patient is seated with the foot dorsiflexed compared to supine with the foot plantarflexed.

03The NordBord device shows high test-retest reliability and moderate correlation with isokinetic dynamometry, despite measuring eccentric rather than isometric force.

STILL TO COME

How it was doneWhat they foundWhat it means for PTsWhat it means for OTsWhat it means for SLPs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

The study did not include a clinician-stabilized HHD condition, so it cannot conclude whether external fixation improved agreement compared to manual stabilization. Foot position was not strictly controlled during the NordBord eccentric testing. The design did not allow for the assessment of test-retest reliability for the externally fixed HHD because body position was counterbalanced across days. Participants were healthy adults, so findings may not apply to clinical populations with pain or injury.