PTOtherPM & R : the journal of injury, function, and rehabilitation2019

A Prospective Assessment of an Adjustable, Immediate Fit, Transtibial Prosthesis.

Timothy Dillingham, Jessica Kenia, Frances Shofer and 1 others

PMID 30734517

WHAT IT FOUND

In 22 people, a new adjustable transtibial prosthesis improved self-reported fit and comfort versus their own devices, lowered overall, anterior tibia, and lateral side peak pressures, but gait speed and step patterns did not differ.

Key findings

01The seven-question fit and comfort score was higher with the IFIT device than with the conventional device (30.9 vs 24.8).

02Gait speed, limp index, stance phase, and double support did not differ significantly between devices.

03No falls or limb ischemia were reported, no mechanical failures occurred, and two participants had minor skin redness or breakdown that healed after adjustments.

STILL TO COME

How it was doneWhat they foundWhat it means for PTs

Read the rest of this summary

You get three full summaries a month, free, and we do not ask for a card. Search, the TL;DRs and your library stay unlimited either way.

Already have one?

What it does not show

There was no separate control group; participants were compared only with their own conventional device. Only 22 of 26 participants completed, and four did not return for follow-up. The trial lasted only two weeks, so it does not show long-term comfort, durability, or skin safety. Gait testing was done in 17 participants; 3 had bilateral transtibial amputations and were not tested, and 2 could not be tested because of equipment repair. Participants were volunteers who may have had less favorable views of their conventional devices and were motivated to try the new device. The pressure film was less accurate for precise pressure values, though it was useful for relative differences. At six months, only 17 participants were contacted, so continued use was not known for all. The device was bulkier for some users, and they needed pants with larger legs. The study tested the whole IFIT prosthesis, including its socket, liner, buckle, and foot. The contribution of any single component cannot be separated. Exclusions included open skin lesions, excessive phantom pain, neurological disorders causing marked weakness or gait impairment, and weight over 260 pounds, so results do not apply to those groups.

Declared interests

One author has a financial interest in the prosthetic system described. He signed a conflict-of-interest management agreement with the University of Pennsylvania Provost for Research. The article is listed as NIH extramural research support.

The easy way to misread this

Do not conclude the IFIT prosthesis is proven superior to conventional prostheses. It was a two-week study of 22 participants compared with their own devices, gait parameters did not differ, and an author has a financial interest in the device.